CI Select
Medizin
Kostenlos · Entwickelt für iPad
For cochlear implant users. Request manual at CISelect.questions@gmail.com or https://med.nyu.edu/research/translational-auditory-research-lab/sites/default/files/ci-select-manual.pdf
NOTE: Instructions for users of Cochlear devices with 22 active electrodes and default frequency allocation table (188-7938Hz). Users with deactivated electrodes or non-default FAT should consult the manual.
1) Audio link to iPhone:
• N7 processor or newer: Use MiniMic 2+ with audio cable between iPhone and Mini-Mic 2+ line-in jack.
• N6 processor or older: Connect iPhone via audio cable to speech processor.
• Deactivate microphone input by:
o Using accessory-only map (consult audiologist)
o Using remote to set 100% accessory/0% microphone mix ratio
NOTE: Current version may not work well with Bluetooth connection between the iPhone and the speech processor. Use wired connection only.
Using CI Select:
• Adjust iPhone volume to comfortable level using speech or continuous audio.
• Select grid boxes (each one represents a different FAT) to compare sound quality and speech intelligibility. Use "LF edge" slider or click boxes to change settings. Red square shows current setting. Default is 188Hz (standard FAT).
• Note: Slight audio delay may occur. If bothersome, minimize acoustic input by removing acoustic component, hearing device in other ear, or using ear plug in opposite ear.
• Test different FATs until finding best intelligibility and sound quality. App makes no permanent changes—share preferred settings with audiologist for permanent map changes.
Grid Searching Tips:
• Tried FATs appear lighter green. Current FAT is red.
• Click "Save Favorite" to mark promising FATs (yellow). Select multiple favorites.
• "Delete last" removes most recent favorite. Repeat to reset grid.
Advanced Option - Extended Frequency Mode: Default grid adjusts LF edge only (sufficient for most users). For full control:
• Click "Extended Frequency Mode" (EFM) below LF edge bar
• This gives access to full grid with LF edges (rows) and HF edges (columns)
• Selected box shows LF & HF intersection
• Toggle EFM switch to return to default view
For non-standard settings, reference the full manual for detailed instructions.
Bewertungen & Rezensionen
Diese App hat noch nicht genügend Bewertungen oder Rezensionen erhalten, um eine Übersicht anzuzeigen.
Startup screen allows user to modify the low-frequency edge of the FAT. An "Extended frequency mode", when selected, allows user to change both low- and high-frequency edges of the FAT.
Das Entwicklungsteam, Mario Svirsky, hat darauf hingewiesen, dass die Datenschutzrichtlinien der App den unten stehenden Umgang mit Daten einschließen können. Weitere Informationen findest du in den Datenschutzrichtlinien des Entwicklungsteams .
Keine Daten erfasst
Der Entwickler erfasst keine Daten von dieser App.
Bedienungshilfen
Der Entwickler hat noch nicht angegeben, welche Bedienungshilfen diese App unterstützt. Weitere Infos
Informationen
- Anbieter
- Mario Svirsky
- Größe
- 27.6 MB
- Kategorie
- Medizin
- Kompatibilität
Erfordert iOS 12.0 oder neuer.
- iPhone
Erfordert iOS 12.0 oder neuer. - iPad
Erfordert iPadOS 12.0 oder neuer. - iPod touch
Erfordert iOS 12.0 oder neuer. - Mac
Erfordert macOS 11.0 (oder neuer) und einen Mac mit Apple M1-Chip (oder neuer). - Apple Vision
Erfordert visionOS 1.0 oder neuer.
- Sprachen
- Englisch
- Altersfreigabe
16+
- 16+
- Häufig
Medizinische Behandlungsinformationen
- Copyright
- © 2005 Mario Svirsky